BCC:
-
Light Fluctuation Method of Measurement
The whole sensor system is consisting of emitter head and receiving head which locate at the opposite side of the flue
In the emitter head, a light-emitting diode with high power is installed and emits light pulse of which wavelength and frequency is fixed.
The light pulse passes through the flue gas and is received by receiving head. When particulate in flue gas passing through infrared light beam,
the intensity of which is directly proportional to the particulate concentration
Operating range: 0-12 m
Ambient temperature: -20 oC ~ 50 oC.
Flue gas temperature: -30 oC ~ +1100 oC.
Measuring range: 0.001~9999 mg/m3
Response time: < 1 sec
Linearity: 0.5 %
Outputs Signal: 4-20 mA/0-10V
- Using the dynamic measuring principle to increase the precision of test result.
- In built flexibility, the concentration of particulate to be monitored feeds back to intelligent control equipment or data recording equipment through the analog output of 4-20mA or 0-10V.
- Unique system of protective gas to ensure clean surface probe.
- Apply to the electricity, metallurgy, cement, high-temperature incineration industries fever.
KSP-I-1000D Particle Analyzer
|